OBJECTIVE:
To lay down the procedure for Operation & Calibration of Gas Chromatography
SCOPE:
This procedure is applicable for the Operation & Calibration of Gas Chromatography
RESPONSIBILITY:
Quality Control Officer / Executive are responsible to follow procedure as per the procedure mentioned in this SOP.
QC Head is responsible for implementation and compliance of the procedure as per this SOP.
ABBREVIATIONS
SOP : Standard Operating Procedure
QC : Quality Control
STP : Standard Test Procedure.
N2 : Nitrogen Gas
HS : Head Space
GC : Gas Chromatography
IPA : Isopropyl Alcohol
ID : Internal Diameter or Identification Code
QCO : Quality Control Officer
PROCEDURE:
Procedure For Operation
Starting The Instrument
Switch on the following:
Gas chromatograph,/Head space sampler (for HS GC) and Data processing unit (Computer).
Open the N2 (carrier gas) valve.
Column Installation
Allow the oven temperature, detector temp, injector temp. to cool below 50°C.
Replace septa and liner if required.
Switch off the oven temperature. Open the door of GC oven.
Make both the column ends flat using cutter if required.
Put the column from injector side while keeping the other end open and purge N2 for a few seconds.
Confirm the N2 flow from the other end by dipping it in 1: 1 mixture of Methanol / IPA and water. Bubbles must form at the outlet.
Insert the other end of column from detector side and ensure tightness at both ends using spanner. Check the leakage from both ends using 1:1 mixture of Methanol / IPA and water and shut the door.
Column Conditioning
Open Zero-Air valve first then the valve of Hydrogen completely.
Optimize flow ratio of zero air and hydrogen, it shall be about 400: 40.Use flow meter to adjust the flow.
Press the ignition button to ignite the flame with the help of lighter.
Put the column on heating with in the maximum temperature limit column till the signal becomes stable.
Method Loading From Software GC Solution
Open Zero-Air valve first then the valve of Hydrogen completely.
Select the option window GC solution. Select instrument No.
Entering the User ID & Password and press OK Key.
Click in the following order FILE / LOAD / METHOD.
To load method choose the path D:\METHOD\PROJECT\METHOD DOWNLOAD Enter Sample Information.
Click at run control, then at sample information to enter following information:
Sample name with Batch No.
Comments W.R.T. Project / Product / Test name
Enter Headspace parameters (if headspace is attached).
After loading various vials for blank, standard and sample into the magazine. Enter the following manually as per the required STP using the front key panel available on the headspace.
Zone temperature
Event time
Vial parameter
STARTING AND STOPPING THE RUN.
Press the start button from headspace key panel or directly from GC key panel (if headspace is not included).
After completion take out the prints of the report generated.
Do the calculation as per formula given in the concerned STP.
In case any problem / clarification contact LAB In-charge.
Calibration of GC
The QCO shall enter the calibration record as per Calibration of Gas Chromatograph (Annexure I).
System Calibration
To check the reproducibility and detectability of the system, perform the following experiments.
Test Column
Fused Silica wide bore capillary column (30 x 0.32mm ID) coated with 0.25µm of 5% Phenyl- 95% polysiloxane (Agilents HP-5 column).
Note: It is Preferable to use one dedicated column for the purpose.
Test Mixture
Prepare sufficient quantity of solution containing about 1.0 mg/ml each of Tetradecane, Pentadecane & Hexadecane in dichloromethane.
Injection volume: Inject 1 µl with 10 µl syringe.
Number of Injection : Six (1 Blank + 5 Test)
GC Parameters
Initial oven temperature : 150°C
Initial time : 0 min
Rate : 10°C / min
Final oven temperature : 200°C
Hold time : 2 min
Flow rate : 2 ml/min.
Injector temperature : 250°C
Detector temperature : 270°C
Carrier flow (N2) : 10 ± 1.0 ml/min
Hydrogen flow : 40 ± 4 ml/min
Zero air flow : 400 ± 40 ml/min
The relative retention times are as follows:
Compound name : Relative retention time
Dichloromethane : ~ 0.19
Tetradecane : ~ 1.00
Pentadecane : ~ 1.20
Hexadecane : ~ 1.50
Set the integration parameters accordingly and proceed with integration and take the prints of the reports generated.
Acceptance Criteria for Calibration.
The percentage relative standard deviation (%RSD) of the ratio of area counts of the peaks corresponding to pentadecane and Hexadecane to that of Tetradecane in the chromatograms of five replicate injections should not be more than 3%.The percentage relative standard deviation (%RSD) of retention times (RT) of the analytes in the five replicate injections should not be more than 2%.
Calibration of Head Space Analyzer
Test Column :
Fused Silica wide bore capillary column (30 x 0.32mm ID) coated with 0.25µm of 5% Phenyl-95% polysiloxane (Agilents HP-5 column)
Test Mixture:
Prepare 0.4%v/v of Ethanol in water (1ml ethanol dilute to 250ml with distilled water).
Test Volume:
Prepare 5 vials each containing 2 ml test solution. Seal the vial with butyl rubber septa and place in the headspace magazine.
GC Parameters:
Oven temperature (Isothermal) : 80°C
Time : 20 min
Injector Temperature : 120ºC
Detector Temperature : 200°C
Carrier Gas : Nitrogen
Carrie Gas Flow : 3.0 ml ± 0.1 ml/ min
Range : 1
Attenuation : 1
Carrier Gas Pressure : 15 psi
Split Ratio : 20:1
Head Space Parameters:
Sample Temperature : 100°C
Needle Temperature : 105ºC
Transfer Line Temperature : 110ºC
Thermostating Time : 22 min
GC Cycle Time : 25 min
Pressurization Time : 3.0 min
Injection Time : 0.04 min
Withdrawal Time : 0.5 min
Acceptance Criteria
For capillary column : The percentage relative standard deviation (%RSD) of the area counts of the peak corresponding to ethanol should not be more than 5.0% (Five replicate injections) and the percentage relative standard deviation (%RSD) of retention times (RT) of analytes in the five replicate injections should not be more than 2.0%
Calibration Frequency : Half Yearly
REMARKS:
If the instrument fails calibration or in the event of a malfunction, report the same to the In-charge, and lodge a complain with the service engineer.
ANNEXURES/REFERENCES :
Annexure I : Calibration of Gas Chromatography.